Experimental Evaluation of a New Electromagnetic Shock Wave Source
1987
Clinical application of new shock wave sources always warrants prior experimental evaluation of the stone disintegration capabilities and characterization of the shock wave impulses. The least possible tissue traumatization is necessary for successful clinical introduction. After the spark-gap generation of shock waves with semi-ellipsoid focussing had been introduced to the urological armamentarium for extracorporeal destruction of renal and ureteral stones in 1980 (Chaussy 1982), new modes of shock wave generation have been developed. In cooperation with the Siemens Company of Erlangen, Germany, a new electromagnetic shock wave source was evaluated by in vitro and in vivo animal experimen…
Assessment of a Modified Acoustic Lens for Electromagnetic Shock Wave Lithotripters in a Swine Model
2013
The acoustic lens of the Modularis electromagnetic shock wave lithotripter (Siemens, Malvern, Pennsylvania) was modified to produce a pressure waveform and focal zone more closely resembling that of the original HM3 device (Dornier Medtech, Wessling, Germany). We assessed the newly designed acoustic lens in vivo in an animal model.Stone fragmentation and tissue injury produced by the original and modified lenses of the Modularis lithotripter were evaluated in a swine model under equivalent acoustic pulse energy (about 45 mJ) at 1 Hz pulse repetition frequency. Stone fragmentation was determined by the weight percent of stone fragments less than 2 mm. To assess tissue injury, shock wave trea…